Problem
When generating a test report in MPE using the ASTM E399 KIC Test Template the report gives multiple warning messages throughout the report (Figure 1):
WARNING! A required variable is not found.
Figure 1 – Warning messages in ASTM E399 KIC Analysis Report
ANSWER
This is an analysis report which requires the creation of an analysis run and report generation in either the Fracture Analyzer or the Fatigue Analyzer application.
From the Tools menu open either the Fracture or Fatique Anayzer Application (Figure 1).
Figure 1 - Fatique Anayzer Application
Right-click on the test run you want to generate the report for and select Anayze Test Run (Figure 2).
Figure 2 – Analyze Test Run selection
Select a test run and click on the Add Analysis button (Figure 3).
Figure 3 – Add Analysis for selected test run
Choose a Select Display option from the drop-down menu and click on OK (Figure 4).
Figure 4 - Select Display and OK
Maximize the Fracture Analyzer window so you can see your Test runs in the left hand column. Right-click on Analysis Run Analysis and select Generate Report (Figure 5).
Figure 5 – Generate Report under the Analysis Run Analysis