Can you add a second HPU board to a FlexTest controller to run a second HPU?

QUESTION

 

Can you add a second hydraulic power unit (HPU) board to a FlexTest controller to run a second HPU?

 

 

ANSWER

 

 

No, although adding a second 493.73 HPU board in the HWI file is possible, you will get the following system load error message when you attempt to load the station (Figure 1):

 

 

“(MakeHydrualicControlTransisitionBoards) Only one HPS transition board allowed per system.  

   Please record this error and call your MTS service representative.”

 

 

cid:image001.png@01D19FAC.51BEB8E0

Figure 1 – System Load Error Window

 

K
Keary is the author of this solution article.

Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.