Yellow Alarm light on Temperature Controller

PROBLEM

 

Yellow Alarm light is lit on 409.83 Temperature Controller (Figure 1).

 

Figure 1 –  409.83 Alarm Indicator Active

 

ANSWER

This could be caused by a possible interlock condition somewhere. Check the Station Manager message-pane window to see if it shows the cause of the interlock (Figure 1). 

Figure 1 – Station Manager Message Pane Window

 

 From pages 28-29 of the attached manual:

cid:image005.png@01D39A97.7FDD8700

Figure 2 – Eurotherm 2216 temperature controller Indicators

 

 

If the temperature controller is not connected to an MTS controller, check that the jumper is installed on P6 connector (Figure 3).   

 

Figure 3 – Temperature Controller P6 Jumper Plug Pinout

 

K
Keary is the author of this solution article.

Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.